2016/8/20,UTC,13:51:36, 2016/8/20,UTC,13:51:36,starting system test... 2016/8/20,UTC,13:51:36,checking limits: limits ok 2016/8/20,UTC,13:51:36,testing power; power ok 2016/8/20,UTC,13:51:36,testing flow; 2016/8/20,UTC,13:51:42,(flow is 9.9 lpm) 2016/8/20,UTC,13:51:47,(flow is 31.6 lpm) 2016/8/20,UTC,13:51:52,(flow is 31.3 lpm) 2016/8/20,UTC,13:51:58,(flow is 31.1 lpm) 2016/8/20,UTC,13:52:3,(flow is 31.0 lpm) 2016/8/20,UTC,13:52:8,(flow is 30.9 lpm) 2016/8/20,UTC,13:52:8,flow ok 2016/8/20,UTC,13:52:8,testing compression; 2016/8/20,UTC,13:52:10,initial pressure 46.2 psia 2016/8/20,UTC,13:52:10,leak checking for 10 seconds... 2016/8/20,UTC,13:52:20,final pressure 45.8 psia 2016/8/20,UTC,13:52:20,yah, compression ok 2016/8/20,UTC,13:52:24,system test succesful 2016/8/20,UTC,13:52:24, 2016/8/20,UTC,13:52:24,AS/SEMIAUTO> starting sample 11... 2016/8/20,UTC,13:52:30,flushing manifold... 2016/8/20,UTC,13:52:30,flushing 5.0 liters with initial measured flow of 9.9 slpm 2016/8/20,UTC,13:52:31,(flow is 31.7 lpm) 2016/8/20,UTC,13:52:34,(flow is 29.6 lpm) 2016/8/20,UTC,13:52:37,(flow is 28.0 lpm) 2016/8/20,UTC,13:52:40,(flow is 27.8 lpm) 2016/8/20,UTC,13:52:41,flushing sample(s)... 2016/8/20,UTC,13:52:41,opening valve 11B... valve open. 2016/8/20,UTC,13:52:44,opening valve 11F... ! valve stuck - couldn't jog it open. 2016/8/20,UTC,13:52:52,closing valve 11F... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:52:53,closing valve 11B... valve closed. 2016/8/20,UTC,13:52:55, 2016/8/20,UTC,13:52:55,! flask valve problem - sample aborted 2016/8/20,UTC,13:52:57,closing valve 11F... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:52:58,closing valve 11B... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:53:1,sample done 2016/8/20,UTC,13:53:5,starting sample 12... 2016/8/20,UTC,13:53:7,flushing manifold... 2016/8/20,UTC,13:53:7,flushing 5.0 liters with initial measured flow of 9.9 slpm 2016/8/20,UTC,13:53:10,(flow is 31.0 lpm) 2016/8/20,UTC,13:53:11,logging at 2016- 8-20 13:53:13 2016/8/20,UTC,13:53:13,(flow is 27.9 lpm) 2016/8/20,UTC,13:53:16,(flow is 27.6 lpm) 2016/8/20,UTC,13:53:18,flushing sample(s)... 2016/8/20,UTC,13:53:18,opening valve 12B... valve open. 2016/8/20,UTC,13:53:21,opening valve 12F... valve open. 2016/8/20,UTC,13:53:24,flushing 10.0 liters with initial measured flow of 29.2 slpm 2016/8/20,UTC,13:53:25,logging at 2016- 8-20 13:53:27 2016/8/20,UTC,13:53:25,(flow is 28.7 lpm) 2016/8/20,UTC,13:53:28,(flow is 28.4 lpm) 2016/8/20,UTC,13:53:31,(flow is 28.4 lpm) 2016/8/20,UTC,13:53:34,(flow is 28.5 lpm) 2016/8/20,UTC,13:53:35,logging at 2016- 8-20 13:53:37 2016/8/20,UTC,13:53:37,(flow is 28.4 lpm) 2016/8/20,UTC,13:53:40,(flow is 28.5 lpm) 2016/8/20,UTC,13:53:43,(flow is 28.4 lpm) 2016/8/20,UTC,13:53:45,logging at 2016- 8-20 13:53:47 2016/8/20,UTC,13:53:46,(flow is 28.5 lpm) 2016/8/20,UTC,13:53:46,pressurizing flask(s)... 2016/8/20,UTC,13:53:46,closing valve 12B... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:53:47,closing valve 12F... valve closed. 2016/8/20,UTC,13:53:49,closing valve 12B... valve closed. 2016/8/20,UTC,13:53:52, 2016/8/20,UTC,13:53:52,! valve problem - sample aborted 2016/8/20,UTC,13:53:54,closing valve 12F... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:53:55,closing valve 12F... ! couldn't start closing; it's stuck or already closed. 2016/8/20,UTC,13:53:55, 2016/8/20,UTC,13:53:55,! valve close problem - sample aborted 2016/8/20,UTC,13:54:0,sample done 2016/8/20,UTC,13:54:0,logging at 2016- 8-20 13:54: 2 2016/8/20,UTC,13:54:0,